| Model Number |
N/A |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during pm, the pressure given out is 10-13 mmhg more than what it was set to.There was no patient involvement.No adverse events were reported as a result of this malfunction.Due diligence is complete and no additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under (b)(4).Telephone number: (b)(6).Review of the most recent repair record identified no related findings/repairs to the reported event.The unit functioned normally.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|