SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP TRIAL INSERT SLOTTED 28/51; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 75000666
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during a thr, one (1) polarcup trial insert slotted 28/51 broke in two pieces.The trial broke when the dr reduced the hip to range the movement and check stability.The dr dislocated the hip and pulled out the broken trial.It was noticed excessive wear and tear on the poly trial.The procedure was resumed, without any delay, using the same device.Patient was not harmed as consequence of this problem.
 
Additional Manufacturer Narrative
H3, h6: it was reported that during a total hip replacement, one (1) polarcup trial insert slotted 28/51 broke in two pieces.The trial broke when the doctor reduced the hip to range the movement and check stability.The doctor dislocated the hip and pulled out the broken trial.It was noticed excessive wear and tear on the poly trial.The procedure was resumed, without any delay, using the same device.Patient was not harmed as consequence of this problem.The device intended for use in treatment was not returned for investigation.A product evaluation was not possible.No batch number was communicated so the document history review was not possible.A complaint history review was performed.The occurrence of the reported failure mode is anticipated with a low risk level as per risk management.Previous investigations demonstrated that the performance of the device is within the risks level that are anticipated in the risk management documentation.No further escalation is required.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.Insufficient information was made available to determine the revision of the instructions for use applicable to the device at the time of manufacturing.However, a review of the current revision was conducted and it revealed that according to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N° 03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.Based on the information provided to the complaint, it is not possible to confirm the reported failure mode.A relationship between the reported event and the device cannot be confirmed, and the root cause of the reported issue remains undetermined.Based on the available information it is not possible to investigate whether the reported device met manufacturing specifications upon release for distribution.Due to insufficient information, it is not possible to speculate about factors which could have contributed to the reported event.There is no need for corrective action because of the limited information provided.Nevertheless, smith + nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP TRIAL INSERT SLOTTED 28/51
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15683501
Report Number9613369-2022-00584
Device Sequence Number6497450
Product Code LXH
UDI-Device Identifier07611996116809
UDI-Public07611996116809
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number75000666
Device Catalogue Number75000666
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date10/27/2022
Supplement Report FDA Received Date11/22/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-