| Model Number |
ADVIA 560 HEMATOLOGY SYSTEM |
| Medical Device Problem Code |
Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An outside of the united states (ous) customer contacted a siemens customer care center (ccc).A siemens customer service engineer (cse) replaced the autoloader for the advia 560 hematology system and the problem did not occur again.No product non-conformance was observed for the autoloader.The cause of the event was limited to the initial autoloader and was resolved by replacing the autoloader.The system is performing according to specifications.No further evaluation of this device is required.
|
| |
|
Event or Problem Description
|
|
The customer reported an issue on their advia 560 hematology system where the autoloader would duplicate the barcode sent to the system from the tube in position 7 to the tube in position 8 of the autosampler rack.The results from the tube in position 8 would therefore be assigned to the incorrect patient.The rest of the tubes were read with the correct barcode.No patient results were reported to the physician(s) for the incorrect patient due to this issue.There are no known reports of patient intervention or adverse health consequences due to this event.
|
| |
|