MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 560 HEMATOLOGY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA 560 HEMATOLOGY SYSTEM Back to Search Results
Model Number ADVIA 560 HEMATOLOGY SYSTEM
Medical Device Problem Code Failure to Read Input Signal (1581)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An outside of the united states (ous) customer contacted a siemens customer care center (ccc).A siemens customer service engineer (cse) replaced the autoloader for the advia 560 hematology system and the problem did not occur again.No product non-conformance was observed for the autoloader.The cause of the event was limited to the initial autoloader and was resolved by replacing the autoloader.The system is performing according to specifications.No further evaluation of this device is required.
 
Event or Problem Description
The customer reported an issue on their advia 560 hematology system where the autoloader would duplicate the barcode sent to the system from the tube in position 7 to the tube in position 8 of the autosampler rack.The results from the tube in position 8 would therefore be assigned to the incorrect patient.The rest of the tubes were read with the correct barcode.No patient results were reported to the physician(s) for the incorrect patient due to this issue.There are no known reports of patient intervention or adverse health consequences due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA 560 HEMATOLOGY SYSTEM
Common Device Name
ADVIA 560 HEMATOLOGY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave.
tarrytown NY 10591
Manufacturer (Section G)
DIATRON MI PIC
registration #: 3009297077
tablas str. 39
budapest, H-109 7
HU   H-1097
Manufacturer Contact
christopher aebig
511 benedict ave.
tarrytown, NY 10591
9144153450
MDR Report Key15685654
Report Number2432235-2022-00292
Device Sequence Number15077321
Product Code GKZ
UDI-Device Identifier00630414601816
UDI-Public00630414601816
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K112755
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberADVIA 560 HEMATOLOGY SYSTEM
Device Catalogue Number11170842
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2022
Initial Report FDA Received Date10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-