• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PLATINUM PLUS; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 39935
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(6).
 
Event or Problem Description
It was reported that tip detachment occurred requiring additional intervention.A 3 018/260 platinum plus guidewire was used in a procedure.However, when pulling back on the wire, the radiopaque tip came off.A snare was used to retrieve the device.The procedure was completed successfully.No patient complications were reported.
 
Additional Manufacturer Narrative
E1: initial reporter city: (b)(6).
 
Event or Problem Description
It was reported that tip detachment occurred requiring additional intervention.A 3 018/260 platinum plus guidewire was used in a procedure.However, when pulling back on the wire, the radiopaque tip came off.A snare was used to retrieve the device.The procedure was completed successfully.No patient complications were reported.It was further reported that the target lesion has complete occlusion, and was located in the non-tortuous and mildly calcified left proximal popliteal artery.Furthermore, a new micro guidewire was used to complete the procedure.After the procedure, there was a significantly improved blood flow in the leg and the patient status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM PLUS
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15685807
Report Number2124215-2022-43687
Device Sequence Number20457287
Product Code DQX
UDI-Device Identifier08714729114574
UDI-Public08714729114574
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K935997
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2025
Device Model Number39935
Device Catalogue Number39935
Device Lot Number0029503677
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer10/31/2022
Initial Report FDA Received Date10/27/2022
Supplement Report FDA Received Date11/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/31/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-