| Model Number |
24-025-49-09 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Seven months after implantation a sternal plate was observed to have fractured in a patient with a history of poor osteogenesis.It was removed and patient was re-plated.
|
| |
|
Additional Manufacturer Narrative
|
|
Multiple mdr reports were filed for this patient.Mdr: 9610905-2022-00010 mdr 9610905-2022-00077.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.During the investigation the product lot number was not identified; therefore, the device history records were not reviewed.Based on the information provided the results conclude that the most likely root cause is patient related.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.Multiple mdr reports were filed for this patient.Mdr: 9610905-2022-00010 mdr 9610905-2022-00077.
|
| |
|