| Model Number |
24-015-29-71 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Lot number: 33312853, 33422518.
|
| |
|
Event or Problem Description
|
|
Twelve months after implant patient presented with new complaint of discomfort and rib plate on a healed rib was found to be fractured on a radiograph.The physician elected to remove the plate per the facility's "standard procedure." no additional hardware was implanted on the healed rib.Cause of plate fracture after consolidation unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed using a stereo microscope revealed tensile cracks.Further observation determined there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at kls se.During the investigation the product lot number identified was reviewed in the device history records.The dhr review showed no discrepancies or anomalies.The investigation results conclude that the root cause for breakages were due to structural failure of the device due to mechanical overload.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.H4 manufacture dates (b)(6) 2019.(b)(6) 2021.
|
| |
|