PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY Back to Search Results
Catalog Number PXSLIMLAN135T90
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a gi bleed in the superior mesenteric artery using a lantern delivery microcatheter (lantern), and a non-penumbra diagnostic catheter.During the procedure, the physician met resistance while advancing the lantern within diagnostic catheter.Next, while retracting the lantern, the physician found that it was broken at the proximal end.Therefore, the diagnostic catheter was removed with the broken piece of the lantern inside.The procedure was completed with a non-penumbra catheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.Placeholder.
 
Additional Manufacturer Narrative
Evaluation of the returned lantern confirmed that the catheter was fractured.It was reported the rhv may have been too tight on the subject lantern.If the rhv is too tight during manipulation of the lantern, it may contribute to increased resistance.Manipulation against this resistance likely contributed to the lantern becoming fractured.Further evaluation revealed ovalizations along the catheter shaft.This damage was incidental to the reported complaint and may have contributed to the reported resistance.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key15686437
Report Number3005168196-2022-00497
Device Sequence Number17960005
Product Code DQY
UDI-Device Identifier00814548016665
UDI-Public814548016665
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K152840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/05/2023
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/13/2023
Device Catalogue NumberPXSLIMLAN135T90
Device Lot NumberF99524
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date10/27/2022
Supplement Report FDA Received Date01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-