• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OPTICROSS 35; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OPTICROSS 35; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number H7493932800350
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that device entrapment occurred.The patient presented with may-thurner syndrome for a venogram.The target lesion was located in a non-tortuous and non-calcified iliac vein.The oc35 imaging catheter was advanced for ultrasound examination of the target lesion.When the device was being removed over a non-boston scientific guidewire, the device became stuck on the wire and the devices had to be removed together through the sheath.The procedure was completed using an alternate device.There were no patient complications.
 
Additional Manufacturer Narrative
Pro code (product code): obj, itx.
 
Additional Manufacturer Narrative
D2b pro code (product code): obj, itx.The device was returned for analysis.Visual inspection revealed the device was returned with a guide wire stuck due residues in the guide wire lumen and with sections of the guide wire lumen torn.Microscopic inspection revealed the transducer housing near the distal tip, also guide wire lumen torn was observed.
 
Event or Problem Description
It was reported that device entrapment occurred.The patient presented with may-thurner syndrome for a venogram.The target lesion was located in a non-tortuous and non-calcified iliac vein.The oc35 imaging catheter was advanced for ultrasound examination of the target lesion.When the device was being removed over a non-boston scientific guidewire, the device became stuck on the wire and the devices had to be removed together through the sheath.The procedure was completed using an alternate device.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS 35
Common Device Name
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15686456
Report Number2124215-2022-44084
Device Sequence Number17976285
Product Code OBJ
UDI-Device Identifier08714729984542
UDI-Public08714729984542
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K200733
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/25/2022
Device Model NumberH7493932800350
Device Catalogue NumberH7493932800350
Device Lot Number0028610965
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/17/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date10/27/2022
Supplement Report FDA Received Date12/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
-
-