• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMENIS LTD VERSAPULSE POWERSUITE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMENIS LTD VERSAPULSE POWERSUITE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0638-803-00
Medical Device Problem Codes Device Displays Incorrect Message (2591); Output below Specifications (3004)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the laser made a noise and then the screen displayed red.Error log showed error 504.The physician was unable to use the laser for the procedure.This event is being reported for aborted/cancelled procedure with a patient under anesthesia or sedation is unknown.
 
Event or Problem Description
It was reported that the laser made a noise and then the screen displayed red.Error log showed error 504.The physician was unable to use the laser for the procedure.This event is being reported for aborted/cancelled procedure with a patient under anesthesia or sedation is unknown.
 
Additional Manufacturer Narrative
Upon receipt of the case report and work order for the console, a thorough review was performed.The allegations were confirmed, as the controller, bricks, and blast shield were damaged.These components could have caused or contributed to the reported event.There were no manufacturing or design issues found; therefore, the cause was traced to component failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAPULSE POWERSUITE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
hakidma 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
LUMENIS LTD
hakidma st 6
yokneam 20692 04
IS   2069204
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key15686485
Report Number2124215-2022-42870
Device Sequence Number20672509
Product Code GEX
UDI-Device Identifier07290109145853
UDI-Public07290109145853
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0638-803-00
Device Catalogue Number0638-803-00
Device Lot Number003-58405
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/13/2022
Supplement Date Received by Manufacturer01/23/2023
Initial Report FDA Received Date10/27/2022
Supplement Report FDA Received Date02/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/06/2007
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-