| Model Number |
380690-03 |
| Medical Device Problem Code |
Optical Discoloration (2999)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint.The fse found memory issue with video slice (vsl) 1, double data rate (ddr) and replaced the vsl 1.The system was tested and verified as ready for use.Isi has requested the vsl 1 involved with this complaint be returned but it has not been received yet.Therefore, the root cause of the alleged customer reported failure mode has not been determined.A follow-up mdr will be submitted upon completion of the failure analysis evaluation and if additional information is received.A review of the site's complaint history identified patient identifier (b)(6) is related record of patient identifier (b)(6) (this complaint).No image or video was available for review.A review of the site's system logs for the reported procedure date was conducted by technical support when the customer called for troubleshooting assistance.Investigation revealed informational error 48266 related to the reported complaint.An event verification confirmed the thymectomy procedure was performed on the reported event date (b)(6) 2022 on system (b)(4).This complaint is considered a reportable malfunction event due to the following conclusion: the customer moved to ssc 2 after the start of the procedure due to color lines in the ssc 1.While there was no harm or injury to the patient, system unavailability after the start of a surgical procedure could likely cause or contribute to an adverse event if it were to recur as it may lead to an injury due to the patient¿s inability to tolerate a procedure change.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted thymectomy surgical procedure, the system had color lines in left eye of the surgeon side console (ssc) 1.The customer completed the procedure with ssc 2.An intuitive surgical, inc.(isi) technical support engineer (tse) reviewed system logs and found informational error 48266.The procedure was completed with no reported injury.Isi followed up with the initial reporter and obtained the following additional information: the reported issue occurred during procedure.The procedure was completed robotically with another surgeon console.There was no patient injury.There were no troubleshooting steps performed during procedure as the customer reported after surgery completion.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the video slice (vsl) 1 involved with this complaint and completed the device evaluation.Failure analysis (fa) could not replicate the customer reported complaint.The vsl was installed and tested on the test system.The system started up without any error, and the image was sharp and clear on both eyes.In-house fa ran 50 power cycles, and all passed.The board remained on the test system for 6 hours in normal operation, while testing other parts, and there is no error logged.The reported issue could not be reproduced.
|
| |
|
Event or Problem Description
|
|
Refer to h10/h11 for follow-up information.
|
| |
|