| Model Number |
TMICL 12.6 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/19/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm tmicl 12.6 implantable collamer lens of -11.50/4.0/111 (sphere/cylinder/axis) was implanted into the patients left eye (os) on (b)(6) 2022.Lens rotation not associated to low vault was observed.Lens rotated 90 degrees.On (b)(6) 2022 the lens was removed and on same day different surgery a spherical lens was implanted of longer length and the problem resolved.Patient will require lasik surgery to address the astigmatism.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation - lens was returned stuck in tape dry.Visual inspection found haptic torn and tape residue throughout the lens.Claim#: (b)(4).
|
| |
|