Model Number DFR00V
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Blurred Vision (2137); Visual Disturbances (2140); Halo (2227)
Date of Event 08/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the patient was experiencing blurry vision and intolerable halos and glare in their right eye (od) resulting in the need to explant the intraocular lens (iol).Distance vision was 20/40 and the patients near was j7.Through follow-up, it was learned that patient could not tolerate the lens was vision is worse.Distance vision is very blurry.The patient's pre-operative (op) uncorrected visual acuity (ucva) was 20/40 distance and best corrected visual acuity (bcva) was j5 near 20/25 distance.Post-operative on (b)(6) 2022 ucva was 20/40-1 distance and j3- near and bcva was 20/40+1 -distance and j2-near.On (b)(6) 2022 ucva was 20/40 distance and j10 near and bcva was 20/25.A non-johnson & johnson lens was implanted as a replacement.There was no incision enlargement, vitrectomy, or sutures were required.There was no patient injury.No additional information was provided.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation? yes.Section d9: returned to manufacturer on: 11/7/2022.Section h3: device evaluated by manufacturer¿ yes.Device evaluation: the complaint lens was received inside of a specimen cup.Visual inspection under magnification revealed that the complaint lens was received cut.The lens was cleaned and, no issues were observed that could contribute to or cause the complaint issues.Based on the return condition of the lens, no further product evaluation could be performed.As a result of the investigation there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, MULTIFOCAL INTRAOCULAR
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15687046
Report Number3012236936-2022-02638
Device Sequence Number6763209
Product Code MFK
UDI-Device Identifier05050474750623
UDI-Public(01)05050474750623(17)241107
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/07/2024
Device Model NumberDFR00V
Device Catalogue NumberDFR00VU175
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer11/30/2022
Initial Report FDA Received Date10/27/2022
Supplement Report FDA Received Date12/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexMale
Patient Weight77 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-