• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENCORE 26 ADVANTAGE KIT; SYRINGE, BALLOON INFLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENCORE 26 ADVANTAGE KIT; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number 2760
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for evaluation.Visual inspection revealed that the gauge needle was at 25 atm when received.Functional test revealed that during preparation, plunger feels hard to activate.A pressure damping test was conducted and the unit passed the test.A vacuum test was conducted and a leakage was observed by the section below the gauge.Side load testing was not able to be performed since leakage was detected.The encore ic gauge accuracy test was not able to be completed since leakage was detected.The device locking mechanism test was not able to be performed since leakage was detected.No other issues were identified during the product analysis.
 
Event or Problem Description
Reportable based on device analysis completed on 11 oct 2022.It was reported that there was an abnormality on the meter of the indeflator.A encore 26 advantage kit was selected for use.During the procedure, it was noted that an abnormality on the meter of the indeflator was observed during balloon catheter inflation.The procedure was completed with another of same device.No patient complications were reported.However, device analysis revealed that the gauge did not return to zero.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE 26 ADVANTAGE KIT
Common Device Name
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15687550
Report Number2124215-2022-42815
Device Sequence Number12467892
Product Code MAV
UDI-Device Identifier08714729180005
UDI-Public08714729180005
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K140745
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/20/2024
Device Model Number2760
Device Catalogue Number2760
Device Lot Number0029075891
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/10/2022
Initial Report FDA Received Date10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-