| Model Number |
2760 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation.Visual inspection revealed that the gauge needle was at 25 atm when received.Functional test revealed that during preparation, plunger feels hard to activate.A pressure damping test was conducted and the unit passed the test.A vacuum test was conducted and a leakage was observed by the section below the gauge.Side load testing was not able to be performed since leakage was detected.The encore ic gauge accuracy test was not able to be completed since leakage was detected.The device locking mechanism test was not able to be performed since leakage was detected.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 11 oct 2022.It was reported that there was an abnormality on the meter of the indeflator.A encore 26 advantage kit was selected for use.During the procedure, it was noted that an abnormality on the meter of the indeflator was observed during balloon catheter inflation.The procedure was completed with another of same device.No patient complications were reported.However, device analysis revealed that the gauge did not return to zero.
|
| |
|