• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM Back to Search Results
Model Number MI1200 SYNCHRONY
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Meningitis (2389)
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received that the device has been explanted.The user had 3 post-operative episodes of meningitis ((b)(6) 2022) for which he had to be hospitalized each time and treated accordingly.It was then decided to perform revision surgery and to remove the device.Reportedly, a ct scan showed an anomaly (common cavity).An transorbital x-ray showed the electrode inside the cochlea.
 
Additional Manufacturer Narrative
The device has been explanted and should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
 
Event or Problem Description
Information was received that the device had been explanted.The user had 3 post-operative episodes of meningitis ((b)(6) 2022) for which he had to be hospitalized each time and treated accordingly.It was then decided to perform revision surgery to remove the device.The user was discharged from the clinic and the user's health improved.
 
Additional Manufacturer Narrative
Conclusion: device investigation did not reveal any device defect or damage, which has been present whilst implanted.Mechanical damages found during investigation are attributable to the removal surgery.According to the information received, the recipient was explanted due to a history of pre- and post-operative recurrent meningitis episodes.Congenital inner ear anomalies, such as the reported common cavity deformity, bear an inherent risk of recurrent meningitis even without a history of otosurgery, as observed in the present case.Further, intra-operative cerebrospinal fluid leakage represents a known risk factor for pneumococcal meningitis.In conclusion, considering the early post-operative onset and the preexisting congenital deformity, the reported event was likely due to a combination of surgical and anatomical factors.A review of the device_s sterilization records shows that the device has been subject to a valid sterilization process, thus no information available points to the implant being the source of infection.However, a device contribution to the subsequent development of the infection cannot be ruled out.This is a final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Common Device Name
COCHLEAR IMPLANT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key15689358
Report Number9710014-2022-00821
Device Sequence Number20672410
Product Code MCM
UDI-Device Identifier09008737094089
UDI-Public(01)09008737094089
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P000025
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/05/2023
Device Model NumberMI1200 SYNCHRONY
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/25/2022
Supplement Date Received by Manufacturer10/25/2022
Initial Report FDA Received Date10/28/2022
Supplement Report FDA Received Date02/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age23 MO
Patient SexUnknown
-
-