GN HEARING A/S GN RESOUND; HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY Back to Search Results
Model Number 21625000
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
(b)(6): on an unknown date charger caught on fire.The hcp informed that patient had no injuries.Patient had wrong plug adaptor attached when brought charger in, but claimed it was plugged in to correct adaptor when the fire started.
 
Additional Manufacturer Narrative
Manufacturer's ref# sf (b)(4).(b)(6): a device history record review have been completed.No deviation or changes were found during manufacturing of the device.Investigation is still in progress, a final report will be submitted upon completion.
 
Event or Problem Description
(b)(6) 2022:.On an unknown date ((b)(6) 2022, best estimate) charger caught on fire.The hcp informed that patient had no injuries.Patient had wrong plug adaptor attached when brought charger in, but claimed it was plugged in to correct adaptor when the fire started.
 
Additional Manufacturer Narrative
Manufacturer's ref# (b)(4).(b)(6) 2022: a device history record review have been completed.No deviation or changes were found during manufacturing of the device.Investigation is still in progress, a final report will be submitted upon completion.(b)(6) 2022: corrected data in g2.Technical investigation concluded: a device history record review have been completed.No deviation or changes were found during manufacturing of the device.Trended case investigation concluded: the usb-c connector has broken down mechanical.It is a known problem with usb connectors, that in rare cases wear or dirt/moist can create a low ohmic path from vbus to gnd, that potential can cause a heating inside of the connector the clinical investigation concluded: a damaged power supply or a power supply of very poor quality could potentially lead to burn injuries and in rare conditions there is a small risk that a wrong charging or discharging or a battery thermal runaway can lead to high temperatures or fire accidents.The chargers have been safety tested according to applicable standards.These risks are reduced as far as possible and the probability of occurrence of these events is considered very low.There are no new clinical aspects (e.G.Unforeseen use or clinical conditions) discovered from this event.Therefore, it is concluded that the current clinical evaluation and risk/benefit conclusions herein are sufficient.Risk register reference: the residual risk after the risk control measure(s) is acceptable.No new hazardous situations revealed and no need to update the risk register the current information indicates need for reporting.Based on the available information, the malfunction may lead to, or might led to, death or a serious injury if it were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GN RESOUND
Common Device Name
HEARING AID, AIR CONDUCTION WITH WIRELESS TECHNOLOGY
Manufacturer (Section D)
GN HEARING A/S
lautrupbjerg 7
ballerup, 2750
DA  2750
Manufacturer Contact
karen frisk
lautrupbjerg 7
ballerup, 2750
DA   2750
MDR Report Key15689988
Report Number3005650109-2022-00061
Device Sequence Number6801968
Product Code OSM
UDI-Device Identifier05708296200061
UDI-Public05708296200061
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number21625000
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/29/2022
Supplement Date Received by Manufacturer09/29/2022
Initial Report FDA Received Date10/28/2022
Supplement Report FDA Received Date11/25/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/17/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-