| Model Number |
E801 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The cause of the event could not be determined.The provided qc data did not exhibit an issue.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable cea elecsys e2g results from the cobas e 801 module.Sample 1 initial result was 23.3 ng/ml.The repeat results were 1.42 ng/ml and 1.42 ng/ml.On (b)(6) 2022, sample 2 initial result was 64.0 ng/ml.The repeat results were 3.15 ng/ml and 3.17 ng/ml.The questionable results were not reported outside of the laboratory.The reagent lot number was 607340 with an expiration date of 30-jun-2023.
|
| |
|