• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR; BIPOLAR SINGLE USE CUTTING LOOP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR; BIPOLAR SINGLE USE CUTTING LOOP Back to Search Results
Model Number 27040GP130-S
Medical Device Problem Codes Material Separation (1562); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer's complaint reference number for this case is (b)(4).Per se robert sapp the doctor used one of the electrodes 27040gp130-s doing a turp procedure.The wire loop portion broke off in the patient.They searched for the loop and were unable to find the loop.Per se the loop was not located during x-ray.
 
Additional Manufacturer Narrative
Per the evaluation findings the loop was broken off and not returned.The orange insulation was melted at the break and is bent in the middle.Additionally, there is black residue on the orange and yellow insulation.The ifu 97000002usca 02-2019, states " the product should not be manipulated or bent.Overloading the instrument by exerting too much force may cause the medical device to break, bend, and malfunction, and consequently injure the patient or user.Do not overload the instruments.Do not bend bent instruments back to their original positions".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUTTING LOOP, BIPOLAR
Common Device Name
BIPOLAR SINGLE USE CUTTING LOOP
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen, gm
Manufacturer (Section G)
KARL STORZ SE & CO.KG
dr.-karl-storz -strasse 34
78532
tuttlingen, gm
Manufacturer Contact
anja fair
2151 e. grand avenue
el segundo, CA 90245
4242188247
MDR Report Key15690452
Report Number9610617-2022-00295
Device Sequence Number6763314
Product Code FAS
UDI-Device Identifier04048551393864
UDI-Public4048551393864
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K061541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number27040GP130-S
Device Catalogue Number27040GP130-S
Device Lot NumberRN11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2022
Initial Report FDA Received Date10/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-