| Model Number |
27040GP130-S |
| Medical Device Problem Codes |
Material Separation (1562); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer's complaint reference number for this case is (b)(4).Per se robert sapp the doctor used one of the electrodes 27040gp130-s doing a turp procedure.The wire loop portion broke off in the patient.They searched for the loop and were unable to find the loop.Per se the loop was not located during x-ray.
|
| |
|
Additional Manufacturer Narrative
|
|
Per the evaluation findings the loop was broken off and not returned.The orange insulation was melted at the break and is bent in the middle.Additionally, there is black residue on the orange and yellow insulation.The ifu 97000002usca 02-2019, states " the product should not be manipulated or bent.Overloading the instrument by exerting too much force may cause the medical device to break, bend, and malfunction, and consequently injure the patient or user.Do not overload the instruments.Do not bend bent instruments back to their original positions".
|
| |
|