Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional codes: nkb, mnh, kwq, kwp device is not distributed in the united states, but is similar to device marketed in the usa.Investigation summary the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that rod ø6 soft l50 ti was observed to have broken at the shaft and deep scratched on the surface.The broken fragment was not received in the evidence provided.A dimensional inspection for the rod ø6 soft l50 tiand was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the rod ø6 soft l50 ti would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Drawing/specifications reviewed: yes, conforms.Dimensional inspection: n/a.Device history.A manufacturing record evaluation cannot be performed due to lot number being unknown.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that on (b)(6) 2022, the surgeon was doing a 1 level instrumented fusion (l5-s1).When screwing the 4th screw in, the screw would not advance further, with 1/4 of the thread remaining.The screw also would not back out.Surgeon tried to use pliers but that did not work.The surgeon was advised to put on the head and short rod, and lock the construct off with sleeve and nuts so that this would turn the screw using screw removal counter torque.This worked, and surgeon removed this set and replaced it with a new set.There was a surgical delay of 10 minutes.This is report 4 of 4 for (b)(4).This report is for a rod.
|