| Model Number |
N/A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
09/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient identifier: requested, not provided.Age & date of birth: requested, not provided.Weight: requested, not provided.Ethnicity: requested, not provided.Race: requested, not provided.Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation: registered technologist.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.Review of the manufacturing history record and the shipping inspection record of the involved product code/lot number combination confirmed that there was not any anomaly in them.A search of the complaint file found no other similar report with the involved product code/lot number combination from other facilities.Terumo medical products (tmp) (importer) registration no.(b)(4) is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.(b)(4).
|
| |
|
Event or Problem Description
|
|
The user facility reported that the catheter involved was inserted in the leg, when it was pulled out, two pieces broke off.Another device was used and was fine.The patient condition was fine.The procedure outcome was successful.The event occurred intra-operative.There was no patient injury/medical or surgical intervention required.There was no equipment or other devices used with the product involved.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the device return date in section d9, update section h3, provide additional information to section b5, and to provide the completed investigation results.Visual inspection of the actual sample upon receipt found no breakage or other anomalies in the appearance over the entire length.Magnifying inspection of the distal end of the actual sample found that the distal tip had been flared.No breakage or missing portion was observed in the distal tip in comparison with a sample from the same product type.Magnifying inspection of the shaft found no deformation or other anomalies in the appearance.Electron microscopic inspection of the flared distal tip found roughness and abrasions on the surface near the flared part.From this, it was inferred that some hard object (e.G., stenotic lesion) was in contact with the involved part.The outer diameter of the actual sample was measured and confirmed to meet the factory's specifications.No anomaly was observed.It was reported that the actual sample was separated into two pieces, however, the investigation results showed that the actual sample had no breakage or missing portion.As one of the possibilities in this case, it was inferred that the actual sample was subjected to pushing force while its distal tip was exposed to and caught on some kind of hard object (stenotic lesion, etc.), which resulted in the flared distal tip.The instructions for use (ifu) states: · "[warnings and precautions] warning: carefully handle the product under fluoroscopy.If any resistance is felt while handling the product, immediately stop the manipulation and find out the cause to avoid damage to blood vessels and separation or breakage of the product." · "[directions for use] - advance the guide wire and the product to the target vessel under fluoroscopy.".
|
| |
|
Event or Problem Description
|
|
Additional information was received on 8 nov 2022: the procedure performed was a runoff percutaneous transluminal angioplasty (pta).The device did not appear damaged prior to use.The pieces broke off inside the patient and were not retrieved.The catheter tip was left inside the patient.The patient was in stable condition.The blood loss was 20cc's.
|
| |
|