| Model Number |
RC2014MIBB |
| Medical Device Problem Code |
Nonstandard Device (1420)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
09/16/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the balloon of a retrograde cardioplegia catheter did not fully expand on one side of the catheter during cardiac bypass.The surgeon had to switch with another device.No patient injury was noted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional narratives the retrograde cannula is used intermittently to deliver retrograde cardioplegia administration during cardiac surgery.If a leak is not noticed during the routine flushing of the device during preparation; exchange to an unaffected device could easily be done during the case since cardioplegia is delivered intermittently and the catheter is easily accessible due to the accessibility of the device and the cardiac structures.The device was returned and product evaluation is ongoing.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional narratives.Updated h3 and h6 per new information received.H3: product evaluation.Customer report of balloon of a rc2014mibb retrograde cardioplegia catheter did not fully expand on one side was confirmed.As received, the balloon was inflated with 5ml of air per ifu without leakage.However, balloon aneurysm was observed.All through lumens were patent without any leakage or occlusion.No visible damage was observed from returned device.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the available information the complaint is able to be confirmed however, a definitive root cause cannot be conclusively determined.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|