• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6534
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that removal difficulties were encountered.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified left anterior descending artery.A 30mm x 2.75mm nc quantum apex balloon catheter was advanced for dilation.The balloon was inflated twice at 28 atmospheres.However, the balloon was very difficult to remove from a non-bsc guidewire although it was in fully deflated state.Eventually, the device was able to be pulled off the guidewire.The procedure was completed successfully with no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of an nc quantum apex balloon catheter with a non-bsc guidewire.The device was visually and microscopically examined.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.The guidewire passed through device with no resistance or issues.The guidewire was inserted through both the rapid port exchange and the tip of the device and encountered no difficulties.A laser micrometer was used to measure the outer diameter of the non-bsc guidewire returned within the balloon catheter.The exposed and undamaged portion of the wire was of compatible use with the balloon.Product analysis did not confirm the reported events as the guidewire passed through the device with no resistance or issues.There was no damage or irregularity found to the device.
 
Event or Problem Description
It was reported that removal difficulties were encountered.The 80% stenosed target lesion was located in the mildly tortuous and mildly calcified left anterior descending artery.A 30mm x 2.75mm nc quantum apex balloon catheter was advanced for dilation.The balloon was inflated twice at 28 atmospheres.However, the balloon was very difficult to remove from a non-bsc guidewire although it was in fully deflated state.Eventually, the device was able to be pulled off the guidewire.The procedure was completed successfully with no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15691726
Report Number2124215-2022-44060
Device Sequence Number15159082
Product Code LOX
UDI-Device Identifier08714729783589
UDI-Public08714729783589
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/14/2024
Device Model Number6534
Device Catalogue Number6534
Device Lot Number0028022467
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/24/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date10/28/2022
Supplement Report FDA Received Date12/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE - RUNTHROUGH BRAND WIRE; GUIDEWIRE - RUNTHROUGH BRAND WIRE
Patient SexUnknown
-
-