• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC EPIDURAL CATHETERIZATION SET; ANESTHESIA CONDUCTION CATHETER Back to Search Results
Model Number IPN040163
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 09/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device investigation is pending and a follow-up report will be issued after the investigation is completed.Teleflex will continue to monitor and trend related events.
 
Event or Problem Description
It was reported that on (b)(6) 2022 in the obstetric block during childbirth, the wingclip snaplock separated from the filter.The consequence was a leak of the treatment and the non-administration of the patient's epidural analgesia.This interruption lasted 1 hour whilst cleaning and replacing the catheter.It cause a risk for the patient.The patient's condition was reported as fine.
 
Additional Manufacturer Narrative
Qn# (b)(4).A device history record review was performed on the flat filter and snaplock assembly with no relevant findings.The customer reported the filter disconnected from the snaplock.The customer returned one flat filter, one snaplock assembly, and an epidural catheter.The filter and snaplock were received connected.The returned components were visually examined with and without magnification.Visual examination of the returned filter and snaplock assembly revealed both components were typical with no observed defects or anomalies.Visual examination of the returned catheter revealed that the catheter appears used as biological material can be seen on the inner coils and adhesive can be seen on the outer extrusion.No other defects or anomalies were observed.Functional inspection was performed on the returned sample.The male lock connector of the returned filter was connected to the returned snaplock assembly and hand tightening the rotating collar to the snaplock assembly with no issues.An attempt to remove the rotating collar from the male lock connector of the returned filter was performed.Using hand pressure, the rotating collar could not be removed even with much effort.In summary, the reported complaint of the filter disconnecting from the snaplock could not be confirmed based on the sample received.A flat filter and snaplock assembly were returned.Using hand pressure, the returned filter's collar could not be removed from the male lock connector even with much effort.Also, there were no connection issues when connecting the returned filter to the returned snaplock assembly.A device history record review was performed on the epidural catheter and snaplock assembly with no relevant findings.The investigation found no evidence to suggest a manufacturing related cause.Based on the functional testing, there was no issues found with the returned sample.No further action is required at this time.
 
Event or Problem Description
It was reported that on (b)(6) 2022 in the obstetric block during childbirth, the wingclip snaplock separated from the filter.The consequence was a leak of the treatment and the non-administration of the patient's epidural analgesia.This interruption lasted 1 hour whilst cleaning and replacing the catheter.It cause a risk for the patient.The patient's condition was reported as fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION SET
Common Device Name
ANESTHESIA CONDUCTION CATHETER
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key15692043
Report Number3006425876-2022-00983
Device Sequence Number6801508
Product Code BSO
UDI-Device Identifier30801902126420
UDI-Public30801902126420
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K140110
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2024
Device Model NumberIPN040163
Device Catalogue NumberJC-05400-E
Device Lot Number71F22F1932
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer11/16/2022
Initial Report FDA Received Date10/28/2022
Supplement Report FDA Received Date11/17/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-