| Model Number |
C502 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter received questionable total protein (tpuc3) assay results from one patient csf sample on a cobas 8000 c 502 module.The initial result tested on the cobas c 502 was not provided and was accompanied by a data flag.The first repeated result tested on the cobas c 502 was 162 mg/l.A new sample was collected and sent to another laboratory to test with a nephelometer, and the results were not provided.Because the nephelometer results do not match the repeated results on the cobas c 502, the initial sample was re-tested on the cobas c 502.The second repeated result on the initial sample was 911 mg/l.It was alleged that the second cobas c 502 repeated results matched the nephelometer results.The reagent lot number was 63953401 and the expiration date was 31-aug-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
A general instrument or reagent problem could not be identified.Based on calibration and qc data a general reagent issue could be excluded.The cause of the event is consistent with sample quality issues.The investigation observed the existence of bubbles/foam on the surface of the sample.
|
| |
|