• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS 8000 C 502 MODULE; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS 8000 C 502 MODULE; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number C502
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.
 
Event or Problem Description
The initial reporter received questionable total protein (tpuc3) assay results from one patient csf sample on a cobas 8000 c 502 module.The initial result tested on the cobas c 502 was not provided and was accompanied by a data flag.The first repeated result tested on the cobas c 502 was 162 mg/l.A new sample was collected and sent to another laboratory to test with a nephelometer, and the results were not provided.Because the nephelometer results do not match the repeated results on the cobas c 502, the initial sample was re-tested on the cobas c 502.The second repeated result on the initial sample was 911 mg/l.It was alleged that the second cobas c 502 repeated results matched the nephelometer results.The reagent lot number was 63953401 and the expiration date was 31-aug-2023.
 
Additional Manufacturer Narrative
A general instrument or reagent problem could not be identified.Based on calibration and qc data a general reagent issue could be excluded.The cause of the event is consistent with sample quality issues.The investigation observed the existence of bubbles/foam on the surface of the sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 8000 C 502 MODULE
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15692137
Report Number1823260-2022-03400
Device Sequence Number7300711
Product Code JJE
UDI-Device Identifier04015630919536
UDI-Public04015630919536
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K100853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC502
Device Catalogue Number05964067001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date10/28/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
-
-