• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72404234
Medical Device Problem Codes Material Puncture/Hole (1504); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an inflatable penile prosthesis (ipp) replacement surgery due to both cylinders ruptured.All components were removed and replaced with new ones.There were no patient complications reported.
 
Event or Problem Description
It was reported that the patient underwent an inflatable penile prosthesis (ipp) replacement surgery due to both cylinders ruptured.All components were removed and replaced with new ones.There were no patient complications reported.
 
Additional Manufacturer Narrative
Upon receipt at our quality assurance laboratory, this inflatable penile prosthesis (ipp) underwent a thorough analysis.The reservoir, cylinders and pump were visually and microscopically examined, and leak tested; the cylinders were also pressure tested and the pump was functionally tested.No anomalies were found with the reservoir.One of the cylinders had broken fabric threads and a bulging in the inner tubing with a hole.The other cylinder passed all testing.The pump did not pass the functional testing as it did not pass the inflation and activation tests based on the information available and analysis results, the bulging and hole identified in the cylinder could have caused or contributed to the reported clinical observation of hole/perforated.Also, the functional anomalies identified in the pump could affect the proper functionality of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Common Device Name
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
alyson harris
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key15692471
Report Number2124215-2022-44243
Device Sequence Number7360016
Product Code FHW
UDI-Device Identifier00878953003474
UDI-Public00878953003474
Combination Product (Y/N)Y
Initial Reporter CountryIT
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date12/15/2019
Device Model Number72404234
Device Catalogue Number72404234
Device Lot Number1000022890
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2022
Supplement Date Received by Manufacturer12/20/2022
Initial Report FDA Received Date10/28/2022
Supplement Report FDA Received Date01/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RESERVOIR:UPN: 720185-01 LOT: 889617005; RESERVOIR:UPN: 720185-01 LOT: 889617005
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-