| Model Number |
N/A |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) the customer has indicated that the product will be returned to zimmer biomet and the investigation is in process.Once the investigation has been completed, a supplemental medwatch 3500a will be submitted.Multiple reports were filed for this event, please see associated reports: 0001526350-2022-00999 and 0001526350-2022-01054.
|
| |
|
Event or Problem Description
|
|
It was reported that the product doesn't hold the pressure and all the air is going out so the product doesn't function.The event timing was during surgery prep.There was no consequences or impact to the patient.Delay is unknown.No further event information available at the time of this report.
|
| |
|
Event or Problem Description
|
|
Due diligence is complete and no additional event information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under: (b)(4).This medwatch is being filed to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.H3 other text: device not returned.
|
| |
|
Event or Problem Description
|
|
Upon receipt of additional information, it has been determined that this device did not cause or contribute to a serious injury or a reportable malfunction.The initial report was forwarded in error and should be voided.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt of additional information, it has been determined that this device did not cause or contribute to a serious injury or a reportable malfunction.The initial report was forwarded in error and should be voided.
|
| |
|
Search Alerts/Recalls
|