Model Number FV410T
Medical Device Problem Codes Mechanical Problem (1384); Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 09/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a progav (part # fv410t) was implanted during a procedure performed on (b)(6) 2017.According to the complainant, the shunt system was believed to be operated in overdrainage and difficult to adjust.The patient underwent a revision procedure on (b)(6) 2022.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 5 years.Height: 104 centimeters (cm).Weight: 14.5 kilograms (kg).Gender: female.
 
Additional Manufacturer Narrative
Investigation: visual inspection: a deformation of the outer housing of the progav valve was observed through the visual inspection.The measurement of the plane parallelism confirmed the visual deformation with a value of -0,064 mm (tolerance 0 ± 0.02 mm).Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the progav was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function is fully operational and the braking force is within the given tolerances.Computer controlled test: to investigate the claim of over-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valve is tested in both the horizontal as well as the vertical positions.The results show that the valve operates within the accepted tolerance in the horizontal position.Internal inspection: after dismantling of the valve, some deposits were found in progav.Result : based on our investigation, we are unable to substantiate the clinical claim of over -drainage.At the time of our investigation, the opening pressure of the valve was within the specified tolerances.However, it is possible that the deposits observed inside the valve could have caused the malfunction in the past.As described in scientific literature, the problem encountered is one of the known, inevitable risks of hc-therapy by shunt implants.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV VALVE
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15692966
Report Number3004721439-2022-00329
Device Sequence Number12602442
Product Code JXG
UDI-Device Identifier04041906132071
UDI-Public4041906132071
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K103003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2021
Device Model NumberFV410T
Device Catalogue NumberFV410T
Device Lot Number20031552
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2022
Initial Report FDA Received Date10/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/01/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age5 YR
Patient SexFemale
Patient Weight14 KG
-
-