The user facility reported that the doctor used the.018 180cm glidewire advantage device while working on an iliac lesion.The wire was put up and taken out, everything was fine.The wire was put back in and it felt stiff; therefore, the wire was pulled out and the tip was gone.The doctor stated he was not being rough with the wire and had been more aggressive with that wire in the past.Access was right femoral through a cordis sheath.The outcome of the procedure was good.The event occurred intra-operative.There was no patient injury/medical or surgical intervention required.Additional information was received on 12 oct 2022: the doctor reported that the tip came off.They retrieved the tip of the wire from the catheter.The patient was in stable condition.Additional information was received on 13 oct 2022: no intervention was necessary.The case was completed successfully.Additional information was received on 14 oct 2022: it was unknown if the patient had any underlying infections and or anti-cancer agents.
|
|
The actual sample upon receipt was a guidewire and its fractured piece.Visual inspection of the actual sample obtained following results.The distal end was approximately 18mm from the distal end.The outer layer had been peeled off, and the wire had been exposed.Approximately 25mm - 130mm from the distal end the outer layer had been peeled off, and the wire had been exposed.The fractured piece was inspected with ft-ir*.As a result, the spectrum was similar to that of the outer layer of the distal end of involved product.Therefore, the fractured piece was the peeled piece of the outer layer at the distal end of actual sample.Ft-ir (fourier transform infrared spectroscopy): an analysis method that irradiates substances with infrared light and measures the transmitted or reflected light for structural analysis and quantitative analysis.Magnifying inspection of the actual sample obtained following results.The distal end was approximately 7mm from the distal end.The gold coil had been unraveled, and the gold coil had been fractured.Peeled piece of the outer layer at the distal end was fractured and the gold coil had been remained.The length of peeled piece was measured and was approximately 17.6 mm.Since the peeled length of outer layer at the distal end of actual sample was approximately 18 mm, it was estimated that the missing was at least 0.4 mm.Approximately 20mm from the distal end the buckling of outer layer had been concentrated on the gold coil side.This outer layer was presumed to be an outer layer that was peeled off at approximately 25mm - 130mm from the distal end.However, it was highly buckling and was unknown if there was missing piece.Approximately 130mm from the distal end: the outer layer had been fractured.Approximately 133mm - 210mm from the distal end.The outer layer had been intermittently flared toward the distal side, and the wire had been exposed.Electron microscopic inspection of the actual sample obtained following results.The side surface of distal end on the wire had been tapered toward the distal side, and the top surface of distal end had traces of processing during manufacturing.No missing part of wire was found due to the same characteristics as the product with the involved product code.Approximately 130mm from the distal end: arc-shaped fracture and wrinkles were found in the outer layer.Approximately 133mm from the distal end: the fracture surface of flared section at the outer layer was partially arc-shaped, and wrinkles were found around it.There were arc-shaped marks on each part, and it was presumed that they were made by abrading against the same object.Buckled section of the outer layer (a): since it was a section where a step occurred between the wire and the gold coil at the boundary of the gold coil, it was presumed that the buckling of outer layer was concentrated.Buckled section of the outer layer (b): the outer layer had been flared toward the distal side.Peeled piece of the outer layer at the distal end (c): the outer layer had been flared, and the gold coil was found.Peeled piece of the outer layer at the distal end (d): a torn shape and wrinkles were found.Since no evidence of elongation was found from the shape of the end of gold coil on the c part, it was presumed to be the gold coil on the distal side.On the contrary, it was presumed that the d part side was the actual sample side.The outer diameter of actual sample (normal section) was measured and confirmed to meet the factory's specifications.From the condition of actual sample, the simulation test was performed assuming that the metal inserter of the combined device came into strong contact when the guidewire was removed.Magnifying inspection of the simulated product found that the outer layer was flared, and the wire was exposed.The condition was similar to that of the actual sample.It was presumed that this event occurred by the following mechanism.Abrasion force was applied while some hard object (e.G., metal device) was in strong contact, and the outer layer was flared toward the distal side.The actual sample was caught on some object (e.G., stenotic lesion or combination device), and the outer layer was buckled and got stuck in the process of pushing and pulling.The pulling force was applied.As a result, the outer layer and the gold coil were fractured.Ashitaka factory manufacturing process has been making efforts in maintaining the quality of the product by performing 100% visual inspection to assure that there is no flaring of the outer layer, exposure of the wire, and anomaly in the appearance.The instructions for use (ifu) has the following warning: "do not manipulate or withdraw the glidewire advantage through a metal entry needle or a metal dilator.Manipulation and/or withdrawal through a metal entry needle or a metal dilator may result in destruction and/or separation of the outer polyurethane coating requiring retrieval.A plastic entry needle is recommended when using this wire for initial placement.".
|