| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device failed self test for defib function.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was duplicated and attributed to the processor/bridge/pace (pbp) board.The pbp board was replaced to resolve the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|