KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR, 24/26FR Back to Search Results
Model Number 27040GP1
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.
 
Event or Problem Description
It was reported that during a hystero resection, the loop broke 3 cm from its end as well as the wire.The ruptured part remained in the endometrium and required removal with hysteroscope forceps.No confirmation of injury or any adverse impact to the patient, user or third party reported.The broken part was retrieved with hysteroscope forceps.Internal karl storz reference number: (b)(4).
 
Additional Manufacturer Narrative
Based on the description of the defect in conjunction with complaint data, the cutting loop has already been used several times and the loop was probably worn.In addition, increased force on the loop may also have caused the sling to break.Therefore, the most probably root cause is mechanical overload due to excessive force.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUTTING LOOP, BIPOLAR, 24/26FR
Common Device Name
CUTTING LOOP
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245-5017
4242188738
MDR Report Key15693894
Report Number9610617-2022-00297
Device Sequence Number17981929
Product Code FAS
UDI-Device Identifier04048551393482
UDI-Public4048551393482
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K061541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number27040GP1
Device Catalogue Number27040GP1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer12/19/2022
Initial Report FDA Received Date10/28/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-