| Model Number |
27040GP1 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hystero resection, the loop broke 3 cm from its end as well as the wire.The ruptured part remained in the endometrium and required removal with hysteroscope forceps.No confirmation of injury or any adverse impact to the patient, user or third party reported.The broken part was retrieved with hysteroscope forceps.Internal karl storz reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the description of the defect in conjunction with complaint data, the cutting loop has already been used several times and the loop was probably worn.In addition, increased force on the loop may also have caused the sling to break.Therefore, the most probably root cause is mechanical overload due to excessive force.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|