MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 8FEN
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Unintended Extubation (4564)
Date of Event 10/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during use, the device had inflated and deflation issue.The cuff inflated and deflated during pretest with the used of syringe with 25 cc of water.Reintubation was required.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the cuff of the device was inflated and deflated during pretest with the used of syringe with 25 cc of water.But, during use, the cuff had inflation and deflation issue.Reintubation was required.
 
Additional Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the received components were one outer cannula, one size 8 inner cannula with integral 15mm twist lock connector (white), one size 8 fenestrated inner cannula with 15 mm connector (green), one size 8 obturator and one cap.The cuff presents a small tear.Functionally, an inflation/deflation test was performed 15cc of air was applied to the cuff using a syringe, and it was observed the cuff deflated after few seconds.It was reported that the cuff had inflation and deflation issue.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: with shiley cuffed models (lpc, fen) the cuff and inflation system should be tested for leakage before inserting the tube.This test can be performed as follows: inflate the cuff with the volume of air indicated in table 1.Then either observe for deflation over several minutes or immerse the tube in sterile saline and observe for air leakage.Deflate the cuff prior to insertion.Under no circumstances should more than 25 mm of mercury air pressure be used to inflate the cuff.Over-inflation of the cuff may cause tracheal damage and may inhibit ventilation.To ease insertion and to guard against cuff perforation from sharp edges of cartilage, the cuff should be tapered back.This can be accomplished by first inflating the cuff.Then gently move the cuff away from the distal tip of the outer cannula towards the swivel neck plate as the residual air is removed by deflation.Do not use any sharp instruments such as forceps or hemostats that would damage the cuff when tapering it.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key15694172
Report Number2936999-2022-00979
Device Sequence Number6821556
Product Code JOH
UDI-Device Identifier60884522007035
UDI-Public60884522007035
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K962173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8FEN
Device Catalogue Number8FEN
Device Lot Number20L00769JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/14/2022
Supplement Date Received by Manufacturer10/14/2022
06/18/2023
Initial Report FDA Received Date10/28/2022
Supplement Report FDA Received Date11/08/2022
07/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-