Model Number UHI-4
Device Problem No Pressure (2994)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/30/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned and customer allegation was confirmed.The issue was due to faulty electropneumatic proportional valve.There was brown colored liquid coming out of the electropneumatic proportional valve via c-tube which happens due to carbon dioxide gas supply to the insufflation unit.Scratches were found on the front panel, power switch, top cover, rear panel and chassis.The investigation is ongoing and a supplemental will be submitted on completion of investigation or if any additional information is available.
 
Event Description
The customer reported to olympus, the high flow insufflation unit was not working and there was no air insufflation.The issue was identified during routine check.There was no patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the following facts obtained from investigation, it is assumed that the reported phenomenon was caused by a failure of the electro-pneumatic proportional valve.The root cause of the failure of the electro-pneumatic proportional valve could not be identified.The root cause of the issue found during inspection of "brown liquid is coming from the electro-pneumatic proportional valve via c-tube¿ has not been identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15694354
MDR Text Key307017552
Report Number3002808148-2022-03662
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/21/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/08/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse