Model Number 21MHPJ-505
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2022, a 21mm sjm masters series hemodynamic plus valve was implanted into a patient.After the valve was implanted, the leaflet on the valve broke.The valve was removed and replaced with 21mm epic supra valve.No additional information was provided.
 
Event or Problem Description
It was reported that on (b)(6) 2022, a 21mm sjm masters series hemodynamic plus valve was implanted into a patient.After the valve was implanted, the leaflet on the valve broke.The valve was removed and replaced with 21mm epic supra valve.Subsequent to the previously filed report, additional information was received.It was reported that when the leaflet was implanted, it was being rotated with rubber shrouded forceps, and it fractured into multiple pieces.All pieces were recovered from patient anatomy.
 
Additional Manufacturer Narrative
An event of valve leaflet fracture during implantation was reported.The device was returned to abbott for investigation and one leaflet was dislodged and fractured while the other leaflet was still in the orifice.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.Based on the information received, the cause of the reported event could not be conclusively determined but may have been caused by some external force being applied to the valve which overstressed the carbon material.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASTERS SERIES EXPANDED HEMODYNAMIC PLUS VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15694781
Report Number2135147-2022-01845
Device Sequence Number6706450
Product Code LWQ
UDI-Device Identifier05414734054201
UDI-Public05414734054201
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/08/2026
Device Model Number21MHPJ-505
Device Catalogue Number21MHPJ-505
Device Lot Number8131908
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date10/29/2022
Supplement Report FDA Received Date11/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-