Model Number 305780
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/03/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bd eclipse¿ needle 25 g x 5/8 in.With detachable 1 ml bd luer-lok¿ syringe experienced a broken barrel.The following information was provided by the initial reporter: broken barrel.
 
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields have been updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 07-nov-2022.H.6.Investigation summary: one photo and one sample were received by our quality team for evaluation.From the photo, the plunger rod was observed to be damaged.The sample was subjected to visual inspection and the plunger rod was observed to be damaged.A review of the internal manufacturing device records and raw material history files for the reported lot number was performed and no recorded quality problems or rejections to this incident were found.The packaging line was reviewed and there are no contact points in the machine that could damage the plunger rod.The 1ml syringe part is supplied from bd canaan and their quality team has completed a supplemental investigation on this.The potential root cause for the damaged plunger defect is associated with the assembly process.It is likely that the plunger thumb rest was mangled as a part of a jam in the plunger feed system and was inadvertently assembled.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.H3 other text : see h10.
 
Event or Problem Description
It was reported that the bd eclipse¿ needle 25 g x 5/8 in.With detachable 1 ml bd luer-lok¿ syringe experienced a broken barrel.The following information was provided by the initial reporter: broken barrel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ECLIPSE¿ NEEDLE 25 G X 5/8 IN. WITH DETACHABLE 1 ML BD LUER-LOK¿ SYRINGE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15694966
Report Number8041187-2022-00629
Device Sequence Number20538571
Product Code FMI
UDI-Device Identifier30382903057802
UDI-Public30382903057802
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K010188
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number305780
Device Catalogue Number305780
Device Lot Number1328213
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date10/29/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-