Model Number HC-1500
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician arrived onsite to inspect the unit and found that the air module was damaged.The air module controls the v40 to allow condensate to release from the sterilizer that collects during a cycle and controls the v46 which is part of the cooling water process to collapse the steam.As the air module was damaged, this resulted in excess condensate to form in the sterilizer which subsequently caused the unit to release excess steam and the reported event to occur.The evolution sterilizer is not under steris maintenance agreement.The user facility is responsible for all maintenance activities.The technician made the necessary repairs, tested the unit, confirmed to be operating according to specifications, and returned to service.A 3-year complaint review indicates this to be an isolated event.No additional issues have been reported.
 
Event or Problem Description
The user facility reported that the drain on their evolution sterilizer was emitting steam.No report of injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUTION STERILIZER
Common Device Name
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key15695855
Report Number3005899764-2022-00053
Device Sequence Number7384753
Product Code FLE
UDI-Device Identifier00724995151539
UDI-Public00724995151539
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHC-1500
Device Catalogue NumberUE03
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/30/2022
Initial Report FDA Received Date10/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/23/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-