Model Number TFGT-A
Medical Device Problem Codes Material Split, Cut or Torn (4008); Intravalvular regurgitation (4068)
Health Effect - Clinical Codes Heart Failure/Congestive Heart Failure (4446); Aortic Valve Insufficiency/ Regurgitation (4450)
Date of Event 10/03/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2019, a 19mm trifecta gt valve was successfully implanted during an aortic valve replacement procedure.In (b)(6) 2022, the valve failure was discovered when the patient presented with symptoms of heart failure and aortic regurgitation.On 3 october 2022, the trifecta gt valve was explanted in a non-emergent procedure, and a 19mm non-abbott device was implanted in the aortic valve.A tear on the explanted valve was noted.The explanting physician did not notice any interaction between any stent post on the original valve and patient anatomy.The patient was reported as stable.
 
Additional Manufacturer Narrative
An event of heart failure, regurgitation and tear on the valve at explant was reported.The investigation found that leaflets 1 and 2 were torn.The was fibrous pannus ingrowth on the inflow surface of leaflets 1 and 2, which would create incomplete coaptation.No inflammation or significant calcifications were present.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.N the absence of any calcification or evidence for infection, the reported event is consistent with a non-calcific leaflet tear.Non-calcific leaflet tear is a form of structural valve deterioration (svd), which is a well-known complication from valve replacement surgery.A non-calcific leaflet tear is commonly attributed to increased operational leaflet stress but may also be related to biological factors which result in tissue degeneration characterized by loss of collagen.Histological evaluation demonstrated mild loss of collagen at one of the tear sites, which could have contributed to the tear.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIFECTA¿ GT VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15695909
Report Number2135147-2022-01850
Device Sequence Number15271872
Product Code LWR
UDI-Device Identifier05415067018205
UDI-Public05415067018205
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P100029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/04/2022
Device Model NumberTFGT-A
Device Catalogue NumberTFGT-19A
Device Lot Number6373304
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer01/12/2023
Initial Report FDA Received Date10/29/2022
Supplement Report FDA Received Date01/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/05/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexUnknown
-
-