U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT5
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
An ophthalmic surgeon reported that during a cataract extraction with intraocular lens (iol) implant procedure, a slightly off-center crack on iol won't have too many consequences found after iol injection.The implant was perfectly normal before going into the injector.The zonula was very fragile, hence did not want to take the risk of removing the iol and putting another one.The surgeon tried to rub the back side of the implant, hoping that it was just dust stuck to the implant, but without success.Additional information was received stating the patient was seen only one day after surgery for the moment, but the visual acuity of the patient at one day after surgery was rather reassuring at 6/10 without correction even if the fissure was visible, a little off center.
 
Additional Manufacturer Narrative
The product was not returned for analysis; the lens remains implanted.Product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints for this lot.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was not returned.Photos were provided of a toric single piece lens in the eye.A mark or material was observed just off center on the right side.Unable to determine the nature or origin of mark.A qualified cartridge was indicated with an unspecified monarch handpiece.The viscoelastic was not provided it is unknown if a qualified product was used.A root cause could not be determined for the reported damage.The lens remains implanted and no issues have been reported.It could not be determined from the provided photos if the pictured area was a crack.Per the instruction for use (ifu): company foldable iols are qualified for use with an company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The use of an unqualified combination may cause damage to the iol and potential complications during the implantation process.The ifu also instructs to completely fill the cartridge with ovd immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15697087
Report Number1119421-2022-02299
Device Sequence Number15299822
Product Code MJP
UDI-Device Identifier00380652251044
UDI-Public00380652251044
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN6AT5
Device Catalogue NumberSN6AT5.225
Device Lot Number15335069
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2022
Supplement Date Received by Manufacturer12/06/2022
Initial Report FDA Received Date10/29/2022
Supplement Report FDA Received Date01/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/25/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III CARTRIDGE D; MONARCH INJECTOR, UNSPECIFIED
Patient Age84 YR