HOLOGIC, INC LOCALIZER; MARKER, RADIOGRAPHIC, IMPLANTABLE Back to Search Results
Model Number HB100
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.We are currently unable to establish a relationship between the device and the issue reported.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported and a definitive root cause for the reported event could not be determined.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event or Problem Description
It was reported that on october 6th, 2022, a lumpectomy procedure was performed and the localizer reader was not functioning properly.The pencil probe from the moment it was plugged in made its audible noise and the numbers on the screen indicating a distance of probe from rfid tag started to flag on the screen and were all over the board (10mm, 49mm, 30mm, etc.) even though the reader and probe were multiple feet from the tag inside the patients breast.The patient was left under general anesthesia for prolonged period of time compared to a traditional wire removal due to equipment/reader malfunctioning and troubleshooting.Surgeon did need to take larger margins than normal for this technology due to the uncertainty of mm distance probe to rfid tag indicated by the localizer reader.The procedure was completed successfully.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCALIZER
Common Device Name
MARKER, RADIOGRAPHIC, IMPLANTABLE
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free zone building b24
san jose 20102
CR   20102
MDR Report Key15697198
Report Number1222780-2022-00341
Device Sequence Number20678630
Product Code NEU
UDI-Device Identifier00855436007037
UDI-Public00855436007037
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K193189
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberHB100
Device Catalogue NumberHB100
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/07/2022
Initial Report FDA Received Date10/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;