W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number RLT351418
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Other relevant history, including preexisting medical conditions: asked but unavailable.(b)(4).W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On (b)(6) 2018, the patient underwent endovascular treatment of an abdominal aortic aneurysm using gore® excluder® aaa endoprostheses.On (b)(6) 2022, the physician made the gore representative aware that the trunk-ipsilateral leg component was reportedly sitting further below the renal arteries than was expected (distance unknown).The physician reported that it was unknown if the device had migrated, if there was aneurysm enlargement, or if there was another cause for the device being more distal than expected.A reintervention will be performed at a later date to place a gore® excluder® aaa aortic extender component.The reintervention has not yet been scheduled.
 
Additional Manufacturer Narrative
B.4.Event description: additional information provided h.6.Investigation findings: code c19 - images were provided and an imaging evaluation was performed.The evaluation determined the following: on the 28-sep-2022 ct, there is contrast outside of the proximal end of the device.This is consistent with a proximal type ia endoleak.The proximal end of the device is ~32 mm distal to the lowest renal.There is one time-point available for evaluation, so migration and aneurysm enlargement cannot be confirmed.Without a pre-treatment ct, possible contributing factors such as amount of device oversizing and condition of the proximal aortic neck at the time of the initial implantation could not be evaluated.H.6.Investigation conclusions: code d12 added.According to the gore® excluder® aaa endoprosthesis instructions for use (ifu), potential device or procedure related adverse events that may occur and/or require intervention or additional intraoperative procedure time include, but are not limited to, component migration and aneurysm enlargement.H.6.Investigation findings: code c21 updated to code c19.H.6.Investigation conclusions: code d16 updated to code d15.
 
Event or Problem Description
On (b)(6), 2022, the reintervention was performed.Two gore® excluder® aaa aortic extender components were used to extend the trunk-ipsilateral leg component proximally.The patient tolerated the reintervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 2 B/P
32470 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
jenna lopez
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key15697251
Report Number3007284313-2022-02178
Device Sequence Number12587844
Product Code MIH
UDI-Device Identifier00733132618491
UDI-Public00733132618491
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/05/2020
Device Model NumberRLT351418
Device Catalogue NumberRLT351418
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer11/14/2022
Initial Report FDA Received Date10/29/2022
Supplement Report FDA Received Date12/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/06/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
-
-