JOHNSON & JOHNSON SURGICAL VISION, INC. FS DISPOSABLE INTERFACE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 590106AN
Medical Device Problem Code Decrease in Suction (1146)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
This report summarizes 240 malfunction events.The events were related to suction loss during laser fire.There were no patient injuries reported associated to the events.
 
Additional Manufacturer Narrative
Lot numbers of the devices and quantity: 60341112 (x 18), unknown (x 14), 60343189 (x 13), 60338314 (x 13), 60357170 (x 12), 60379006 (x 10), 60332109 (x 10), 60371231 (x 9), 60343419 (x 9), 60327645 (x 8), 60375374 (x 7), 60347544 (x 7), 60333690 (x 6), 60377740 (x 6), 60358902 (x 6), 60347300 (x 5), 60376065 (x 5), 60357624 (x 5), 60358903 (x 5), 60347543 (x 5), 60292889 (x 4), 60368822 (x 4), 60319240 (x 4), 60358905 (x 4), 60336893 (x 4), 60377741 (x 3), 60343420 (x 3), 60336894 (x 3), 60330263 (x 3), 60341113 (x 3), 60388199 (x 2), 60287678 (x 2), 60319241 (x 2), 60324362 (x 2), 60323882 (x 2), 60327630 (x 2), 60330262 (x 2), 60324359 (x 2), 60332108 (x 1), 60381549 (x 1), 60297881 (x 1), 60301370 (x 1), 60385030 (x 1), 60303339 (x 1), 60355767 (x 1), 60294306 (x 1), 60325739 (x 1), 60320254 (x 1), 60383021 (x 1), 60323225 (x 1), 60317514 (x 1), 60276236 (x 1), 60377739 (x 1), 60334442 (x 1).One hundred and five (105) investigations were completed during the period.No product deficiency was identified.A review of the records related to the device that included labeling, manuals, trending, and risk documentation reviews was performed.A review of the device history record (dhr) showed that the system and its components met all specifications prior to being released.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.The patient interface (pi) suction ring may lose suction during a procedure.Label copy states corneal fixation vacuum loss can occur.There are several factors that may contribute to suction issues such as doctor¿s technique in applying the suction ring to the cornea, doctor¿s technique in squeezing the pi clip to secure the suction ring to the pi cone and patient anatomy affecting the interface between the patient¿s cornea and the suction ring.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Device evaluation: seventy investigations were completed during the period and the findings were as follow: two devices had operational problems - one of them a misshaped and bent suction ring edges and when functionally tested it did not met specification.The second one failed the functional tested as the syringe would not stay depressed and vacuum pressure gradually decreased.One device was returned incomplete and testing could not be performed.Sixty seven devices were found within specification.A review of the records related to the device that included labeling, manuals, trending, and risk documentation reviews was performed.A review of the device history record (dhr) showed that the system and its components met all specifications prior to being released.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.
 
Additional Manufacturer Narrative
Correction - it was initially reported there were 240 events however, it was found there were 242 events.This also changed quantity for the unknown lot number events from 14 to 16 as both events were for unknown lot numbers.
 
Additional Manufacturer Narrative
Device evaluation: two (2) investigations were completed during the period.No product deficiency was identified.For one of the devices it arrived broken.A review of the records related to the device that included labeling, manuals, trending, and risk documentation reviews was performed.A review of the device history record (dhr) showed that the system and its components met all specifications prior to being released.The risks and mitigations associated with the complaint issue are identified in existing risk documents and no new risks were identified as part of this investigation.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FS DISPOSABLE INTERFACE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
31 technology drive
irvine CA 92618
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15697614
Report Number3012236936-2022-02720
Device Sequence Number17819172
Product Code HNO
UDI-Device Identifier05050474535305
UDI-Public(01)05050474535305
Combination Product (Y/N)N
PMA/510(K) Number
K060372
Number of Events Summarized240
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 04/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number590106AN
Device Catalogue Number590106AN
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/30/2022
Supplement Date Received by Manufacturer01/27/2023
01/27/2023
04/28/2023
Initial Report FDA Received Date10/30/2022
Supplement Report FDA Received Date01/27/2023
01/27/2023
04/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-