• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ TORIC SINGLEPIECE IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number SN6AT7
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lens was returned positioned incorrectly posterior side up in the lens case.Solution was dried on the lens.Both haptics are bent in the gusset and distal area.The optic was torn/split/cracked and cut into to multiple pieces, typical of insertion and removal.Multiple scratches, smaller cracks, and scrape marks were observed on the anterior and posterior sides of the lens, which may have been interpreted as the reported "indention" complaint.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.The file indicated the use of a qualified cartridge and handpiece.The associated viscoelastic was not provided.It is unknown if a qualified product was used.Upon return, the lens was observed to be placed into the lens case incorrectly resulting in damage.Lens damage was observed that may have been interpreted as the reported "indention" complaint.Other lens damage was observed that may have occurred due to how the lens was positioned in the lens case for return.All lenses are 100% inspected for cosmetic attributes and the damage exhibited by the returned complaint sample would not have met company release criteria.Based on our observation, and the review of manufacturing records, it can be reasonably concluded that the damage is not manufacturing related.Due to the presence of surgical solution and the condition of the returned sample, the damage is most likely related to customer handling.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A nurse reported during the intraocular lens implantation, it was noticed that the lens had indentations on it.The surgery was competed on the same day.Additional information has been requested and received stating the lens was inserted into the eye and then cut out when the "indention" was seen on the lens.No harm done to the patient, another lens was implanted, and the surgery was successful.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ TORIC SINGLEPIECE IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15697867
Report Number1119421-2022-02308
Device Sequence Number17981848
Product Code MJP
UDI-Device Identifier00380652275767
UDI-Public00380652275767
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 10/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSN6AT7
Device Catalogue NumberSN6AT7.120
Device Lot Number15329953
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2022
Initial Report FDA Received Date10/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III CARTRIDGE D; MONARCH III INJECTOR
Patient Age62 YR
Patient SexMale
-
-