• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JABIL CIRCUIT (SHANGHAI) LTD LATITUDE COMMUNICATOR; EXTERNAL COMMUNICATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JABIL CIRCUIT (SHANGHAI) LTD LATITUDE COMMUNICATOR; EXTERNAL COMMUNICATOR Back to Search Results
Model Number 6290
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the communicator showed no lights, and the communicator power cord was emitting sparks.A boston scientific company representative advised not to continue troubleshooting due to sparking of the communicator power cord.The company representative advised to replace the communicator power cord.No adverse patient effects were reported.The communicator remains in service.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory additional information from product investigation indicates that the allegation against the communicator was confirmed.The returned communicator output voltage was out of specification.There was no measurable evidence of electrical overstress in the communicator modem circuitry.The communicator plastic case was removed, and a burnt smoky smell was observed.The communicator behavior was consistent with a high-power electrical overstress event that was attributed to the environment.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the communicator showed no lights, and the communicator power cord was emitting sparks.A boston scientific company representative advised not to continue troubleshooting due to sparking of the communicator power cord.The company representative advised to replace the communicator power cord.No adverse patient effects were reported.The communicator remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LATITUDE COMMUNICATOR
Common Device Name
EXTERNAL COMMUNICATOR
Manufacturer (Section D)
JABIL CIRCUIT (SHANGHAI) LTD
600 tian lin rd
shanghai
CH 
Manufacturer (Section G)
JABIL CIRCUIT (SHANGHAI) LTD
600 tian lin rd
shanghai 20023 3
CH   200233
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15698227
Report Number2124215-2022-44378
Device Sequence Number12165616
Product Code LWS
UDI-Device Identifier00802526544224
UDI-Public00802526544224
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P910077/S140
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number6290
Device Catalogue Number6290
Device Lot Number5400699
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/18/2022
Supplement Date Received by Manufacturer02/06/2023
Initial Report FDA Received Date10/31/2022
Supplement Report FDA Received Date02/08/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age88 YR
Patient SexMale
-
-