• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EIT EMERGING IMPLANT TECHNOLOGIES GMBH IMPLANT INSERTER; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EIT EMERGING IMPLANT TECHNOLOGIES GMBH IMPLANT INSERTER; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number LET30100
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in germany as follows: it was reported that on (b)(6) 2022, the cage did not move from the inserter after implant placement.Surgery was delayed by 30 minutes as a result of this event.The surgeon removed the cage with the inserter and used a cage from a different company.The procedure was then completed successfully.No further information is available.This report involves one implant inserter.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h4 the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Part # let30100 lot # e19di1312 supplier: eit ¿ batch1: lot qty of 240 units were released on 06 may 2020 with no discrepancies.No ncrs were generated during production.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the cage llif hi h 14mm 0deg 45/18, was stuck at the tip of the inner shaft of the implant inserter, p/n: let30100.The complaint device exhibits signs of normal use.A dimensional inspection was unable to be performed due to not being applicable to the complaint condition.A functional test to disassemble the devices, was conducted.It was discovered that the cage was stuck at the tip of the inner shaft.It was not possible to turn the knob to loosen the inner shaft, therefore, the devices could not be disassembled.The complaint condition was replicated.Since it was not possible to disassemble the device, a root cause for this failure mode can not be established.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the implant inserter, p/n: let30100 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANT INSERTER
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
EIT EMERGING IMPLANT TECHNOLOGIES GMBH
eisenbahnstrasse 84
wurmlingen D-785 73
Manufacturer (Section G)
EIT EMERGING IMPLANT TECHNOLOGIES GMBH
Manufacturer Contact
daniel seng
eisenbahnstrasse 84
wurmlingen D-785-73
GM   D-78573
6103142063
MDR Report Key15698396
Report Number3013730328-2022-00082
Device Sequence Number6352791
Product Code MAX
UDI-Device Identifier04260557837733
UDI-Public04260557837733
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K181644
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLET30100
Device Catalogue NumberLET30100
Device Lot NumberE19DI1312
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/07/2022
Supplement Date Received by Manufacturer12/08/2022
Initial Report FDA Received Date10/31/2022
Supplement Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/06/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONDUIT LLIF LUMEN H 14MM 0DEG 45/18
Patient SexUnknown
-
-