MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPICC HF CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS) STYLET (5F); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPICC HF CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS) STYLET (5F); CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of regq0757 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported: lines seem to be leaking.Bed was found saturated after infusion, leaking at site.Additional information: location: right mid-thigh.Length: 50 cm.Yes this was used on a patient, no harm was done.Treated for fever, n/v, covid positive.Stat lock securement device.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history review, applicable previous investigation(s), sample (if available), applicable manufacture records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter is unconfirmed as the failure could not be reproduced.One 5fr t/l powerpicc catheter was returned for evaluation.An initial visual observation showed use residues and an intentional cut at the 50 cm marking of the powerpicc catheter.The three clamps were observed to be disengaged.Functional testing revealed all three lumens were patent to infusion with no observed leaks.No leaks were observed during pressurization of the sample.A microscopic observation revealed the catheter had been cut with a sharp instrument at the 50 cm depth marker.No observed splits or holes were identified on any location of the catheter, therefore, the complaint of a leak was unconfirmed.
 
Event or Problem Description
It was reported: lines seem to be leaking.Bed was found saturated after infusion, leaking at site.Additional information: location: right mid-thigh femoral.Length: 50cm.Yes this was used on a patient, no harm was done.Treated for fever, n/v, covid positive.Stat lock securement device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPICC HF CATHETER WITH SHERLOCK 3CG TIP POSITIONING SYSTEM (TPS) STYLET (5F)
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15698569
Report Number3006260740-2022-04188
Device Sequence Number18260983
Product Code LJS
UDI-Device Identifier00801741143762
UDI-Public(01)00801741143762
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K091324
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberN/A
Device Catalogue NumberCK000633B
Device Lot NumberREGQ0757
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/07/2022
Supplement Date Received by Manufacturer01/10/2023
Initial Report FDA Received Date10/31/2022
Supplement Report FDA Received Date01/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age29 YR
Patient SexMale
Patient Weight100 KG
-
-