| Model Number |
XR30V |
| Medical Device Problem Code |
Failure to Form Staple (2579)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unknown procedure, while using the linear stapler the cartridge misfired there was no b formation, so another gun was used to complete the surgery there were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 11/15/2022.D4: batch # 479a77.Investigation summary: the product was for evaluation.Visual inspection was conducted on the returned reload.Visual analysis of the returned sample revealed that the xr30v reload was received fully fired and with the right side damaged.No functional test was performed due to the condition of the reload.The device was tested for functionality with a test reload and it fired and formed all the staples as intended.The device fired without any difficulties and the staples meet the staple form release criteria.It is possible that the reload was not loaded correctly.To reload the device the tracks on each side of the reload should be used as guides to align the reload within the jaws of the instrument.When the reload is properly aligned, push the reload into the instrument until it is fully seated, and a click is heard.If the instructions are not following the reload may became damage.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances related to the reported complaint condition were identified.
|
| |
|
Search Alerts/Recalls
|