| Model Number |
PC9000 |
| Medical Device Problem Codes |
Image Display Error/Artifact (1304); Optical Problem (3001)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/04/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: "loss of image quality likely due to damage in the camera port connection." probable root cause/s: reported failure mode was confirmed and root cause is due to a faulty fvp board.The device was returned for investigation and the reported failure mode was reproduced.The reported failure mode will be monitored for future re occurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image.
|
| |
|
Additional Manufacturer Narrative
|
|
Correcting fda registration number from 0002936485 - endoscopy (san jose) to 3012345110 - endoscopy (novadaq).
|
| |
|