Catalog Number 03666794001
Medical Device Problem Code Image Display Error/Artifact (1304)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
The initial reporter stated they had a display issue with a coaguchek xs meter.Upon completion of a display check, segments were missing from the results field of the device.No patients have been tested with the meter since the customer noticed the missing segments.
 
Additional Manufacturer Narrative
The meter was requested for investigation.
 
Additional Manufacturer Narrative
The returned meter was investigated.The meter could not be turned on.The device data and fault memory cannot be read out.Investigation of the returned meter found that the battery contacts and the circuit board are contaminated by liquid (leaked battery).This affects the battery contacts on the circuit board, which are interrupted by this, and the contacts of an electronic component.The root cause of the event was contamination of the contacts due to improper handling or maintenance by the user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK ® XS SYSTEM
Common Device Name
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15700489
Report Number1823260-2022-03445
Device Sequence Number12823604
Product Code GJS
UDI-Device Identifier00365702126107
UDI-Public00365702126107
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K062925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03666794001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer11/18/2022
Initial Report FDA Received Date10/31/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-