Model Number 0220220300
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/10/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there was a thermal event.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
Alleged failure: customer states safelight fell off scope and contributed to drape burn.[update - devices were not warm/hot to the touch.No devices were smoking.No patient/user injuries, only drape was damaged and replaced.] the failure alleged in the complaint record was not confirmed/duplicated during the product investigation.The probable root cause: not following the user guide safety/warnings.Advise to calibrate the light source unit.Before operating this device, please read this operating manual thoroughly and carefully.When using a light source, fire and/or severe injury may result to the patient, user or inanimate objects if the instructions in this manual are not followed.All light sources can generate significant amounts of heat (exceeding 41 °c/106 °f) at the scope tip, the scope light post, the light cable tip, and/or near the light cable adapter.Higher levels of brightness result in higher levels of heat.Adjust the brightness level of the camera and the monitor to the minimum brightness necessary to adequately illuminate the surgical site.In addition, adjust the internal shutter of the camera higher in order to run the light source at a lower intensity.Avoid touching the scope tip or the light cable tip to the patient, and never place them on top of the patient, as doing so may result in burns to the patient or user.In addition, never place the scope tip, the scope light post, the light cable adapter, or the light cable tip on the surgical drapes or other flammable material, as doing so may result in fire.Always place the light source in standby mode and disable env mode before the scope is removed from the light cable or the device is unattended.The scope tip, scope light post, light cable adapter, and light cable tip will take several minutes to cool off after being placed in standby mode, and therefore may still result in fire or burns to the patient, user, or inanimate objects.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
 
Event or Problem Description
It was reported that there was a thermal event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, L10 LED LIGHT SOURCE WITH AIM
Common Device Name
CONFOCAL OPTICAL IMAGING
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
benjamin ly
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15700950
Report Number0002936485-2022-00621
Device Sequence Number20881163
Product Code OWN
UDI-Device Identifier07613327051377
UDI-Public07613327051377
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K142310
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0220220300
Device Catalogue Number0220220300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/10/2022
Supplement Date Received by Manufacturer10/10/2022
Initial Report FDA Received Date10/31/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/01/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-