• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. SL-PLUS AND SLR-PLUS; HIP, STEM, CLP, STANDARD, SIZE#5, IMPLANT, POLISHED NECK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. SL-PLUS AND SLR-PLUS; HIP, STEM, CLP, STANDARD, SIZE#5, IMPLANT, POLISHED NECK Back to Search Results
Model Number 426-10-050
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 10/12/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - patient had a hip fracture.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as hip fracture.The previous surgery and the surgery detailed in this event occurred 1.9 years apart.The in vivo time for the stem is 2 years, from the primary surgery.Initial or prolonged hospitalization was required.The healthcare professional indicated there was a serious risk to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to hip fracture.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL-PLUS AND SLR-PLUS
Common Device Name
HIP, STEM, CLP, STANDARD, SIZE#5, IMPLANT, POLISHED NECK
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445
Manufacturer Contact
james mcmahon
9800 metric blvd
austin, tx 78758-5445 
MDR Report Key15701941
Report Number1644408-2022-01438
Device Sequence Number12817369
Product Code JDI
UDI-Device Identifier00888912079532
UDI-Public(01)00888912079532
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K932481
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/12/2026
Device Model Number426-10-050
Device Catalogue Number426-10-050
Device Lot Number405N1062
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/12/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date10/31/2022
Supplement Report FDA Received Date11/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
400-04-360 LOT 883B1198; 400-05-300 LOT 889B1074
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient Age75 YR
Patient SexFemale
-
-