|
HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
| Model Number |
1650 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that three batteries exhibited power switching, the patient noted the batteries beeping and disconnecting.The batteries have been replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Additional products: brand name: heartware ventricular assist system ¿ battery, model #: 1650, catalog #: 1650, expiration date: 30-apr-2021, serial or lot#: (b)(4), udi #: (b)(4), device available for evaluation: no, mfg date: 21-apr-2020, labeled for single use: no, (b)(4); brand name: heartware ventricular assist system ¿ battery, model #: 1650, catalog #: 1650, expiration date: 30-apr-2021, serial or lot#: (b)(4), udi #: (b)(4), device available for evaluation: no, mfg date: 21-apr-2020, labeled for single use: no, (b)(4).If additional information is received, the event will be updated and a supplemental report will be sent.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report is being submitted for device evaluation.Product event summary: three (3) batteries (b)(6) were returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the devices in relation to the reported event.Failure analysis of the returned batteries revealed that the devices passed visual inspection and functional testing.The batteries were able to adequately provide power to a test controller with no anomalies observed.Log file analysis was not performed since controller log files were not available for analysis.As a result, the reported event could not be confirmed.The batteries were lubricated prior to release.Applicable risk documentation and experience with events of similar circumstances were considered; a possible root cause of the reported premature power switching and ¿disconnection¿ events can be attributed, but not limited, to communication errors and/or momentary disconnections due to temporary corrosion of the controller-port/power-source pins.However, the reported beeps are most likely attributed to momentary disconnections between the controller and battery.Additional products: d4: serial or lot#: (b)(6) d9: yes, return date: 09-nov-2022 h3: yes dev rtn to mfr? yes h6: fda method code(s): b01 h6: fda results code(s): c19 h6: fda conclusion code(s): d14 d4: serial or lot#: (b)(6) d9: yes, return date: 09-nov-2022 h3: yes dev rtn to mfr? yes h6: fda method code(s): b01 h6: fda results code(s): c19 h6: fda conclusion code(s): d14 investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|
|
|