• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL GALAXY G3 4MM X 12CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL GALAXY G3 4MM X 12CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number GLY120412
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2021
Type of Reportable Event Malfunction
Event or Problem Description
The healthcare professional reported that during an endovascular embolization procedure that took place on (b)(4) 2021, the complaint coil, a 4mm x 12cm galaxy g3 coil (gly120412 / 30494398) was used with the concomitant excelsior® sl-10® microcatheter (stryker) but it was not able to be detached.The complaint coil was removed from the patient.It was not known if continuous flush was maintained.There was no report of any patient adverse event or patient complication.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).Procode: procode is krd/hcg.[conclusion]: the healthcare professional reported that during an endovascular embolization procedure that took place on 04-nov-2021, the complaint coil, a 4mm x 12cm galaxy g3 coil (gly120412 / 30494398) was used with the concomitant excelsior® sl-10® microcatheter (stryker) but it was not able to be detached.The complaint coil was removed from the patient.It was not known if continuous flush was maintained.There was no report of any patient adverse event or patient complication.Based on complaint information, the device is not available to be returned for analysis.A review of manufacturing documentation associated with this lot (30494398) presented no issues during the manufacturing or inspection processes related to the reported complaint.There were no nonconformances related to device manufacture or inspection.Product analysis cannot be conducted as the product was not returned for analysis.No determination of causes and possible contributing factors could be made.As such, the investigation will be closed.Failure to detach is a known potential issue associated with the use of this device.The instructions for use (ifu) contains several precautions related to this issue and includes instructions for troubleshooting the situation should it be encountered during use.With the information available and without the complaint product available to be returned for analysis, the reported issue in the complaint cannot be confirmed.Based on the manufacturing documentation review, there is no indication that the event is related to the device manufacturing process.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint sample; however, it is possible that clinical and procedural factors, including device manipulation / interaction may have contributed to the reported failure.As part of the post market surveillance program, information from this complaint is trended for statistical signals and corrective / preventive action may be triggered at a later time.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
Manufacturer¿s ref.No: (b)(4).The purpose of this mdr submission is to include the additional event information received on 16-nov-2022.[additional information]: on 16-nov-2022, additional information was received.The information indicated that the complaint coil was successfully removed from the patient.It was not stretched when it was removed; the complaint coil was still attached to the delivery system when it was removed.A pre-deployment electrical check was performed.The fault light was not seen during the procedure.Upon pressing the power button, all lights did illuminate; the green system ready light did illuminate.The audible signal beep was heard during the detachment cycle.All connections fit without the application of force.There was no clinically significant delay in the procedure as a result of the reported issue.E.1: initial reporter phone: (b)(6).The manufacturer will submit a supplemental report if new facts arise which materially alter information submitted in a previous mdr report.Additional information will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY G3 4MM X 12CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CODMAN & SHURTLEFF, INC. (FREMONT)
47709 fremont blvd
fremont CA 94538
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
303552-6892
MDR Report Key15704355
Report Number3008114965-2022-00664
Device Sequence Number6584380
Product Code KRD
UDI-Device Identifier10886704077480
UDI-Public10886704077480
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model NumberGLY120412
Device Catalogue NumberGLY120412
Device Lot Number30494398
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2022
Supplement Date Received by Manufacturer11/17/2022
Initial Report FDA Received Date10/31/2022
Supplement Report FDA Received Date11/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
EXCELSIOR® SL-10® MICROCATHETER (STRYKER)
Patient SexUnknown
-
-